- Home >> News >> Research Progress
Research Progress
Evaluation of the Uncertainties Caused by the Forward Scattering in Turbidity Measurement of the Coagulation Rate
Date:2010-07-21Source:
The forward scattering light (FSL) received by the detector can cause uncertainties in turbidity measurement of the coagulation rate of colloidal dispersion, and this effect becomes more significant for large particles. In this study, the effect of FSL is investigated on the basis of calculations using the T-matrix method, an exact technique for the computation of nonspherical scattering. The theoretical formulation and relevant numerical implementation for predicting the contribution of FSL in the turbidity measurement is presented. To quantitatively estimate the degree of the influence of FSL, an influence ratio comparing the contribution of FSL to the pure transmitted light in the turbidity measurement is introduced. The influence ratios evaluated under various parametric conditions and the relevant analyses provide a guideline for properly choosing particle size, measuring wavelength to minimize the effect of FSL in turbidity measurement of coagulation rate.
This paper was published as:
Xu SH, Sun ZW. Evaluation of the Uncertainties Caused by the Forward Scattering in Turbidity Measurement of the Coagulation Rate. LANGMUIR, 26(10):6908-6918(2010)